GoldGreen Dental Slow-Speed Air Motor Reversible

Upgrade your dental toolkit with the GoldGreen Dental Slow-Speed Air Motor Reversible, engineered for general dental procedures such as polishing, finishing, and cavity preparation. Designed with precision and user comfort in mind, this handpiece delivers smooth operation, reliable performance, and optimal cooling—ensuring both patient comfort and practitioner efficiency.

Key Features

  • Versatile Clinical Use – Perfect for polishing, finishing, and cavity preparation in various dental procedures.
  • Ergonomic Design – Lightweight, comfortable grip minimizes fatigue during extended use.
  • Smooth & Quiet Operation – Delivers consistent torque and stable performance at low noise levels (≤70 dB).
  • Effective Cooling System – External cooling keeps the working area at optimal temperature, enhancing patient comfort.
  • Durable Construction – Built to withstand repeated sterilization and daily clinical demands.

Safety Instructions

  • Operate only by trained dental professionals.
  • Always sterilize before initial use and between patients.
  • Disconnect handpiece from air supply before disassembly or maintenance.
  • Use only burs specified for this model to prevent injury or malfunction.

System Components

  • 1 × Slow-Speed Air Motor Reversible

Assembly Instructions

  1. Align the handpiece with the air motor connection.
  2. Rotate clockwise until firmly attached.

Maintenance & Lubrication

A. Lubrication Before Sterilization

  1. Detach the handpiece from the tubing.
  2. Spray handpiece oil into the rear coupling until oil exits the head.
  3. Operate the handpiece for 30 seconds to distribute lubricant.
  4. Wipe off excess oil.

B. Frequency

  • After each use and prior to sterilization.
  • Daily, if not in use, to prevent internal drying.

Sterilization Guidelines

  • Autoclave at 134°C (273°F) for 4 minutes.
  • Use only approved pouches for sterilization.
  • Do not exceed 135°C.
  • Allow full cooling before next use.
  • Do not soak or submerge in disinfectants.

Technical Specifications

Specification Value
Gear Ratio 1:1 (Direct Drive)
Air Pressure 245–392 kPa
Water Pressure 198 kPa (2 kgf/cm²)
Rotation Speed 20,000 – 27,000 rpm
Bur Applicable 2.334 – 2.355 mm
Noise Level ≤70 dB
Cooling System External